TENDRIL STS
MDR report key: 4191981
·
Received January 13, 2014
Report
- Report Number
- 2017865-2014-04642
- Event Type
- Malfunction
- Date Received
- January 13, 2014
- Date of Event
- May 8, 2012
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- DTB
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION DESCRIPTION: THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE. THE LEAD WAS OTHERWISE NORMAL.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT, THERE WAS A LOSS OF VENTRICULAR CAPTURE. THE PACEMAKER WAS REPLACED, BUT THE ISSUE WAS STILL SEEN. THE ISSUE WAS RESOLVED BY EXPLANTING AND REPLACING THE LEAD.