FDA Adverse Event
Malfunction
Summary report: N
ZEPHYR XL DR
MDR report key: 4191961
·
Received January 13, 2014
Report
- Report Number
- 2017865-2014-04548
- Event Type
- Malfunction
- Date Received
- January 13, 2014
- Date of Event
- November 21, 2011
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- LWP
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION DESCRIPTION: LEAD FIT APPROPRIATELY IN DEVICE CONNECTOR HEADER AND SETSCREW FUNCTIONED NORMALLY. COMPLAINT COULD NOT BE CONFIRMED. DEVICE EXHIBITED NORMAL OUTPUT CHARACTERISTICS.
Description of Event or Problem · 1
IT WAS REPORTED THAT WHEN THE LEAD WAS INSERTED AND SETSCREW SCREWED DOWN THE PULSE GENERATOR EXHIBITED NO CAPTURE. THE PATIENT IS DEPENDENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28404 | ZEPHYR XL DR | IMPLANTABLE PACEMAKER PULSE GENERATOR | LWP | ST. JUDE MEDICAL, INC., CRMD | 5820 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 84 YR |