FDA Adverse Event Injury Summary report: N

IDENTITY ADX XL DR

MDR report key: 4191742 · Received January 13, 2014

Report

Report Number
2017865-2014-04784
Event Type
Injury
Date Received
January 13, 2014
Date of Event
January 17, 2012
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
LWP
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. THE REPORTED FIELD EVENT OF NOISE WAS NOT CONFIRMED. ELECTRICAL TESTS REVEALED NORMAL DEVICE CHARACTERISTICS. NO ANOMALY WAS FOUND.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED NOISE. THE DEVICE WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29793 IDENTITY ADX XL DR IMPLANTABLE PACEMAKER PULSE GENERATOR, LWP LWP ST. JUDE MEDICAL, INC., CRMD 5386

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention 1788TC/52 BAM21979