MDR report key: 4191728 · Received January 13, 2014

Report

Report Number
2017865-2014-04579
Event Type
Injury
Date Received
January 13, 2014
Date of Event
February 9, 2012
Manufacturer
ST. JUDE MEDICAL, CRMD
Product Code
MXC
PMA / PMN Number
K081365
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION: ANALYSIS WAS NORMAL. NO ANOMALY WAS NOTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING IMPLANT OF THE IMPLANTABLE CARDIAC MONITOR THE PHYSICIAN SAW A SPARK WHILE USING ELECTROCAUTERY. SUBSEQUENTLY, THE DEVICE EXHIBITED BACKUP VVI OPERATION. THE DEVICE WAS NOT USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33707 CONFIRM IMPLANTABLE CARDIAC MONITOR MXC ST. JUDE MEDICAL, CRMD DM2100

Patients

Seq Age Sex Outcome Treatment
1 54 YR Required Intervention