FDA Adverse Event Malfunction Summary report: N

HOMECHOICE PRO

MDR report key: 4191475 · Received October 22, 2014

Report

Report Number
1416980-2014-36857
Event Type
Malfunction
Date Received
October 22, 2014
Date of Event
August 20, 2014
Report Date
September 26, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. A VISUAL INSPECTION WAS PERFORMED. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. A REVIEW OF THE ALARM LOG IDENTIFIED THE OCCURRENCE OF THE INCREASED INTRA-PERITONEAL VOLUME EVENT. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN #5) ALARM WAS IDENTIFIED IN THE LOG. THIS ALARM INDICATES THAT THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME. THE ALARM OCCURRED ON (B)(6) 2014 AT 03:24:53. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
673179 HOMECHOICE PRO SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 67 YR