HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-36857
- Event Type
- Malfunction
- Date Received
- October 22, 2014
- Date of Event
- August 20, 2014
- Report Date
- September 26, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. A VISUAL INSPECTION WAS PERFORMED. THE HOMECHOICE DEVICE RECEIVED A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDED FUNCTIONAL AND ELECTRICAL TESTING OF THE DEVICE. A REVIEW OF THE ALARM LOG IDENTIFIED THE OCCURRENCE OF THE INCREASED INTRA-PERITONEAL VOLUME EVENT. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, A HIGH DRAIN ERROR 105 (NIGHT DRAIN #5) ALARM WAS IDENTIFIED IN THE LOG. THIS ALARM INDICATES THAT THE PATIENT DRAINED GREATER THAN 200% OF THE MAXIMUM PRESCRIBED CYCLE FILL VOLUME. THE ALARM OCCURRED ON (B)(6) 2014 AT 03:24:53. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 673179 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |