FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4191406 · Received October 21, 2014

Report

Report Number
2531779-2014-29881
Event Type
Malfunction
Date Received
October 21, 2014
Report Date
October 8, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 12/22/2014 - DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/03/2014 WITH THE FOLLOWING FINDINGS: ON (B)(4) 2014, THE PUMP'S BLACK BOX RECORDED A SINGLE LOSS OF PRIME WARNING DUE TO A LOW, NON-ZERO FORCE. THE PUMP WAS EXERCISED FOR 24 HOURS AT A RATE OF 1 UNIT/HOUR WITH NO LOSS OF PRIME WARNINGS OR OTHER ALARMS. THE FORCE SENSOR WAS FOUND TO BE CALIBRATED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THERE WERE MULTIPLE LOSS OF PRIME WARNINGS WITH DIFFERENT CARTRIDGES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
669445 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 7 YR