MITEK VAPR COOLPULSE 90 ELECTRODE
Report
- Report Number
- 1221934-2014-00450
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- DEPUY MITEK
- Product Code
- GEI
- PMA / PMN Number
- K100638
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE COMPLAINT DEVICE WAS NOT RECEIVED. HOWEVER, THE PICTURE PROVIDED CONFIRMS THAT THE METAL TIP IS BROKEN AND SEPARATED FROM THE SHAFT. WITHOUT PHYSICALLY EXAMINING THE DEVICE, A ROOT CAUSE FOR THE EVENT CANNOT BE DETERMINED. THIS TIP BREAK FAILURE ON CP90 ELECTRODES ARE BEING INVESTIGATED UNDER A SUPPLIER CAPA. FOUR POSSIBLE ROOT CAUSES HAVE BEEN IDENTIFIED RELATED TO DESIGN, MANUFACTURING AND USE. CORRECTIVE ACTIONS WHICH INVOLVE DESIGN AND PROCESS IMPROVEMENT ACTIVITIES WERE IDENTIFIED FOR EACH OF THE ROOT CAUSE AND ARE BEING CURRENTLY INVESTIGATED. A BATCH RECORD REVIEW HAS BEEN CONDUCTED AND OUR RESULTS INDICATE THAT THIS BATCH OF PRODUCT WAS PROCESSED WITHOUT INCIDENT AND THEREFORE THERE IS NO INTERNALLY ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM. A REVIEW INTO THE COMPLAINTS SYSTEM REVEALED ONE OTHER SIMILAR COMPLAINT FOR THIS LOT OF DEVICES THAT WERE RELEASED TO DISTRIBUTION. (B)(4). MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD.
LOOSE METALPART FROM THE TIP OF THE VAPR ELECTRODE. **ADDITIONAL INFORMATION FROM THE AFFILIATE WHAT TYPE OF PROCEDURE WAS THIS? DEKOMPRESSION SHOULDER. WHAT WAS THE DATE OF THE PROCEDURE? (B)(6) 2014. DID THE TIP FALL INTO THE PATIENT? YES. HOW WAS IT REMOVED? WITH A GRASPER. HOW WAS THE PROCEDURE COMPLETED? HOW LONG DID THIS FAILURE EXTEND THE PROCEDURE TIME, MORE THAN 30 MINUTES? NO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670223 | MITEK VAPR COOLPULSE 90 ELECTRODE | ELECTROSURGICAL: CUTTING AND COAGULATING | GEI | DEPUY MITEK | NA | U1404069 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |