FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 4191038
·
Received October 21, 2014
Report
- Report Number
- 2032227-2014-41236
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Date of Event
- September 22, 2014
- Report Date
- September 22, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED DURING TESTING. THE DEVICE HAD MINOR SCRATCHED LCD WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT AT BATTERY TUBE THREADS AREA, AND CRACKED RESERVOIR TUBE LIP.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 669933 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-722LWWS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |