COULTER LH 780 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2014-02623
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Date of Event
- September 25, 2014
- Report Date
- September 25, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FSE FOUND A CLOGGED PORT IN WASTE CHAMBER VC26 CAUSING THE WASTE CHAMBER TO OVERFLOW. THE FSE CLEARED THE CLOG AND THE INSTRUMENT RAN WITHOUT LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).
THE CUSTOMER REPORTED AN UNCONTAINED CLEAR FLUID LEAK OF APPROXIMATELY TWENTY (20) ML UNDER THE TRIPLE TRANSDUCER MODULE (TTM) INSIDE THE COULTER LH 780 HEMATOLOGY ANALYZER. THERE WAS NO REPORT OF INJURIES, ERRONEOUS RESULTS, DIRECT PHYSICAL CONTACT WITH THE LEAK, OR CHANGE TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES, GOGGLES, AND LABORATORY COAT WHEN THE LEAK OCCURRED. A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670124 | COULTER LH 780 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |