FDA Adverse Event Malfunction Summary report: N

COULTER LH 780 HEMATOLOGY ANALYZER

MDR report key: 4191030 · Received October 21, 2014

Report

Report Number
1061932-2014-02623
Event Type
Malfunction
Date Received
October 21, 2014
Date of Event
September 25, 2014
Report Date
September 25, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FSE FOUND A CLOGGED PORT IN WASTE CHAMBER VC26 CAUSING THE WASTE CHAMBER TO OVERFLOW. THE FSE CLEARED THE CLOG AND THE INSTRUMENT RAN WITHOUT LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN UNCONTAINED CLEAR FLUID LEAK OF APPROXIMATELY TWENTY (20) ML UNDER THE TRIPLE TRANSDUCER MODULE (TTM) INSIDE THE COULTER LH 780 HEMATOLOGY ANALYZER. THERE WAS NO REPORT OF INJURIES, ERRONEOUS RESULTS, DIRECT PHYSICAL CONTACT WITH THE LEAK, OR CHANGE TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT CONSISTING OF GLOVES, GOGGLES, AND LABORATORY COAT WHEN THE LEAK OCCURRED. A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER SITE TO EVALUATE THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
670124 COULTER LH 780 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA