PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-08922
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Date of Event
- July 15, 2014
- Report Date
- July 17, 2014
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS REVEALED THAT BROKEN RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION. THE INSULIN PUMP PASSED PRIME, DISPLACEMENT, BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY ALARM TESTS. NO MOTOR ERROR ALARMS NOTED AND THE MOTOR TESTED OK.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING MOTOR ERROR AND NO DELIVERY WHILE PRIMING. DURING TROUBLESHOOTING, CUSTOMER STATED THAT THEY DO NOT USE THE SENSOR FEATURE AND WAS ABLE TO REWIND THE INSULIN PUMP. CUSTOMER STATED THEY DO NOT RECALL ANY SIGNIFICANT EVENTS THAT LEAD TO THE ALARM OR IF THE DEVICE WAS DROPPED. ADVISED CUSTOMER TO DISCONTINUE USE OF DEVICE AND REVERT TO BACK UP PLAN. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 250 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.