FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4191012 · Received October 21, 2014

Report

Report Number
2032227-2014-41228
Event Type
Malfunction
Date Received
October 21, 2014
Date of Event
May 15, 2014
Report Date
September 22, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP PASSED DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33AND EXCESSIVE NO DELIVERY TESTS. NO UNEXPECTED EXCESSIVE NO DELIVERY ALARM NOTED. THE INSULIN PUMP HAS CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, MISSING END CAP STICKER AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE BUTTONS ON THE INSULIN PUMP ARE STICKING. IT WAS ALSO REPORTED THAT VIBRATION ON THE INSULIN PUMP IS VERY LOW. ADVISED INSULIN PUMP WOULD BE REPLACED. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
669914 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAS

Patients

Seq Age Sex Outcome Treatment
1 65 YR