FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4190990 · Received October 21, 2014

Report

Report Number
2032227-2014-41197
Event Type
Malfunction
Date Received
October 21, 2014
Date of Event
September 22, 2014
Report Date
September 22, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP BUTTONS WERE UNRESPONSIVE. CUSTOMER CHANGED THE BATTERY AND IT DID NOT WORK. NO BLOOD GLUCOSE READING PROVIDED. CUSTOMER HAD TO DISCONTINUE TO USE THE INSULIN PUMP AND FOLLOWED MEDICAL PRESCRIPTION FOR INSULIN SHOTS UNTIL REPLACEMENT ARRIVES. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668932 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1