FDA Adverse Event
Malfunction
Summary report: N
G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM
MDR report key: 4190989
·
Received October 21, 2014
Report
- Report Number
- 3004753838-2014-31291
- Event Type
- Malfunction
- Date Received
- October 21, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 25, 2014
- Manufacturer
- DEXCOM, INC.
- Product Code
- MDS
- PMA / PMN Number
- P120005
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE COMPLAINT DEVICE WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY INSPECTED AND NO DEFECT WAS FOUND. FUNCTIONAL TESTING WAS PERFORMED AND THE REPORTED FAULT COULD NOT BE REPRODUCED AND THERE WAS NO FAILURE DETECTED. THE DEVICE WAS DETERMINED TO BE OPERATING WITHIN THE REQUIRED SPECIFICATIONS WITHOUT MALFUNCTION. HOWEVER, BASED ON THE DATA LOG, THE COMPLAINT OF LOSS OF ANTENNA WAS CONFIRMED. A CAPA HAS BEEN INITIATED FOR THIS ISSUE.
Description of Event or Problem · 1
PATIENT'S MOTHER CONTACTED DEXCOM TECHNICAL SUPPORT ON (B)(6) 2014 TO REPORT AN OUT OF RANGE SIGNAL ON (B)(6) 2014. PATIENT'S MOTHER DID NOT REPORT ANY INJURY OR MEDICAL INTERVENTION AT THE TIME OF CONTACT WITH DEXCOM TECHNICAL SUPPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 670097 | G4 PLATINUM PEDIATRIC CONTINUOUS GLUCOSE MONITORING SYSTEM | MDS | MDS | DEXCOM, INC. | MT22430 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 10 YR |