THV 1000-29 3F AORTIC BIO 29MM
Report
- Report Number
- 3010152718-2014-00006
- Event Type
- Injury
- Date Received
- October 21, 2014
- Date of Event
- September 23, 2014
- Report Date
- December 1, 2014
- Manufacturer
- MEDTRONIC 3F THERAPEUTICS, INC.
- Product Code
- LWR
- PMA / PMN Number
- P060025
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
CONCLUSION: THE DEVICE HISTORY RECORD WAS REVIEWED AND SHOWED THAT THIS PRODUCT MET ALL MANUFACTURING SPECIFICATIONS FOR PRODUCT RELEASED FOR DISTRIBUTION. NO ISSUES WERE IDENTIFIED THAT WOULD HAVE IMPACTED THIS EVENT. BASED ON THE ANALYSIS, A CONCLUSIVE CAUSE OF THE PERFORATION OF THE LEAFLET COULD NOT BE DETERMINED. THE RECEIPT CONDITION OF THE VALVE MAY HAVE ALTERED THE GROSS OBSERVATIONS. DISTORTION OF THE VALVE MAY HAVE CREATED UNEVEN LEAFLET MOVEMENT THEREBY CONTRIBUTING TO THE TEAR ON THE LEAFLET. THE PERFORATION DOES NOT APPEAR TO HAVE BEEN INDUCED DURING THE MANUFACTURING PROCEDURE, BASED ON THE DEVICE HISTORY RECORD REVIEW. THE ROOT CAUSE FOR THE REPORTED CLINICAL OBSERVATION COULD NOT BE DETERMINED. MEDTRONIC WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS.
UPON RECEIPT AT MEDTRONIC¿S QUALITY LABORATORY, THE VALVE WAS RECEIVED DISTORTED IN AN OVAL SHAPE. THE VALVE TISSUE WAS DESICCATED. ALL LEAFLETS WERE STIFF BUT SLIGHTLY FLEXIBLE, EXCEPT WHERE THE TISSUE WAS DESICCATED. DISTORTION IN THE VALVE MAY HAVE CONTRIBUTED TO A PROLAPSED APPEARANCE IN LEAFLETS 1 AND 2. A LARGE PERFORATION IN THE BELLY OF LEAFLET 2 MAY BE ASSOCIATED WITH UNEVEN LEAFLET MOVEMENT THAT MAY HAVE RESULTED IN A HINGE POINT. THE TOPS OF ALL COMMISSURES HAD BEEN REMOVED DURING THE EXPLANT. A TRACE OF PANNUS REMAINED ATTACHED TO LEAFLET 3 ADJACENT TO THE 3-2 COMMISSURE. A REMNANT OF PANNUS ON THE INFLOW EXTENDED ONTO THE TUNICA OF LEAFLET 1FROM THE SEWING RING. A REMNANT OF PANNUS REMAINED ATTACHED TO THE OUTFLOW EDGE OF THE SEWING RING ADJACENT TO LEAFLET 2, EXTENDING TO THE OUTFLOW MARGIN OF ATTACHMENT. RADIOGRAPHY SHOWED NO EVIDENCE OF MINERALIZATION IN THE VALVE.
IT WAS REPORTED THAT THE DEVICE WILL BE RETURNED FOR ANALYSIS. A SUPPLEMENTAL REPORT WILL BE FILED WHEN THE ANALYSIS/INVESTIGATION IS COMPLETED. (B)(4).
MEDTRONIC RECEIVED INFORMATION THAT THIS AORTIC BIOPROSTHETIC VALVE WAS EXPLANTED 18 MONTHS AFTER ITS IMPLANT DUE TO SEVERE INSUFFICIENCY. IT WAS REPORTED THAT THERE WERE OBSERVATIONS OF A VALVE LEAFLET PERFORATION AND A SMALL DETACHMENT OF THE BASE OF THE VALVE FROM THE ANNULUS. THE VALVE WAS REPLACED WITH ANOTHER MANUFACTURER¿S VALVE WITH NO FURTHER ADVERSE PATIENT EFFECTS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 666877 | THV 1000-29 3F AORTIC BIO 29MM | HEART-VALVE, NON-ALLOGRAFT TISSUE | LWR | MEDTRONIC 3F THERAPEUTICS, INC. | 1000-29 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00075 YR | Required Intervention |