FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4190439 · Received October 21, 2014

Report

Report Number
2032227-2014-41117
Event Type
Malfunction
Date Received
October 21, 2014
Date of Event
September 22, 2014
Report Date
September 22, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH AN INTERMITTENT BUTTON RESPONSE, DUE TO CORRODED KEYPAD TRACES. NO RAMPING NUMBERS ANOMALY NOTED. INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

THIS MDR RELATED TO THE (B)(4) MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4). CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED SHE WAS ATTEMPTING TO GIVE A BOLUS FOR LUNCH THE NUMBERS KEPT RAMPING UP AND WOULD NOT STOP. CUSTOMER STATED SHE TOOK THE BATTERY OUT OF THE DEVICE AND THEN NONE OF THE BUTTONS WERE WORKING. CUSTOMER'S BLOOD GLUCOSE WAS 119 MG/DL. CUSTOMER DOES NOT RECALL ANY SIGNIFICANT EVENT THAT MAY HAVE CAUSED THE KEYPAD ISSUE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
667543 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 68 YR