FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4190274 · Received October 21, 2014

Report

Report Number
2531779-2014-29800
Event Type
Malfunction
Date Received
October 21, 2014
Report Date
September 1, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED FOR INVESTIGATION AND DEVICE EVALUATION WAS COMPLETED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP CASE WAS REMOVED AND CORROSION WAS FOUND UNDERNEATH THE INTERNAL BATTERY COMPARTMENT IS CRACKED BELOW FINGER PAD THE BLACK BOX CONFIRMED TIME AND DATE RESET TO DEFAULT AFTER BATTERY CHANGE, PUMP WAS EXERCISE FOR 24 HOURS, THEN TOOK OUT THE BATTERY AND THE PUMP WAS EXERCISE FOR 6 HOURS AND THE TIME AN DATE WAS RESET TO DEFAULT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THE PUMP WAS RETURNED FOR PRODUCT ANALYSIS INVESTIGATION AND THE EVALUATION REVEALED A CRACKED BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON THE RESULTS OF AN INVESTIGATION COMPLETED ON (B)(4) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668918 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 9 YR