FDA Adverse Event Malfunction Summary report: N

HOMECHOICE

MDR report key: 4190155 · Received October 21, 2014

Report

Report Number
1416980-2014-36765
Event Type
Malfunction
Date Received
October 21, 2014
Date of Event
August 29, 2014
Report Date
September 26, 2014
Manufacturer
BAXTER HEALTHCARE - LARGO
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED AND THE EVALUATION IS COMPLETE. THIS IS AN ANCILLARY SERVICE EVENT. DURING THE EVENT HISTORY LOG REVIEW, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. UPON CONCLUSION OF THE INVESTIGATION, THE CAUSE OF THE EVENT WAS DETERMINED TO BE USE ERROR, TIDAL TOTAL ULTRAFILTRATION REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014, AT 23:55:18. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1560ML, INDICATING THE HOME PATIENT DRAINED 1365ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 2000ML. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668610 HOMECHOICE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - LARGO

Patients

Seq Age Sex Outcome Treatment
1