FDA Adverse Event Malfunction Summary report: N

TRUERESULT

MDR report key: 4190125 · Received September 18, 2014

Report

Report Number
1052693-2014-00342
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 23, 2014
Report Date
December 23, 2015
Manufacturer
NIPRO DIAGNOSTICS, INC.
Product Code
NBW
PMA / PMN Number
K080641
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USE HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.

Additional Manufacturer Narrative · 1

PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).

Description of Event or Problem · 1

CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS AROUND 85-90 MG/DL. RESULTS IN MEMORY WERE 56 MG/DL AND 47 MG/DL. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (47) AND THE NORMAL RESULT (90) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS AROUND 85-90MG/DL. RESULTS IN MEMORY WERE 56MG/DL AND 47MG/DL. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (47) AND THE NORMAL RESULT (90) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
577693 TRUERESULT BLOOD GLUCOSE SYSTEM NBW NIPRO DIAGNOSTICS, INC. TRUERESULT PP1745

Patients

Seq Age Sex Outcome Treatment
1 0 YR