TRUERESULT
Report
- Report Number
- 1052693-2014-00342
- Event Type
- Malfunction
- Date Received
- September 18, 2014
- Date of Event
- August 23, 2014
- Report Date
- December 23, 2015
- Manufacturer
- NIPRO DIAGNOSTICS, INC.
- Product Code
- NBW
- PMA / PMN Number
- K080641
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). MOST LIKELY UNDERLYING ROOT CAUSE OF MALFUNCTION USE HAD AN INACCURATE REFERENCE. INVESTIGATION COMPLETED AND PRODUCT EVALUATED WITH NO DEFECTS FOUND.
PRODUCT NOT YET RETURNED FOR EVALUATION. (B)(4).
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS AROUND 85-90 MG/DL. RESULTS IN MEMORY WERE 56 MG/DL AND 47 MG/DL. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (47) AND THE NORMAL RESULT (90) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.
CONSUMER COMPLAINT OF LOW BLOOD RESULTS. CALLER NORMALLY IS AROUND 85-90MG/DL. RESULTS IN MEMORY WERE 56MG/DL AND 47MG/DL. BASED ON PARKES ERROR GRID ANALYSIS, THE BIAS BETWEEN THE LOWEST RESULT GIVEN (47) AND THE NORMAL RESULT (90) IS LOCATED IN ZONE B/C. NO ADVERSE EVENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 577693 | TRUERESULT | BLOOD GLUCOSE SYSTEM | NBW | NIPRO DIAGNOSTICS, INC. | TRUERESULT | PP1745 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |