FDA Adverse Event Malfunction Summary report: N

PLMA DVC V11.51 1 R

MDR report key: 4190118 · Received September 18, 2014

Report

Report Number
9615050-2014-05312
Event Type
Malfunction
Date Received
September 18, 2014
Date of Event
August 21, 2014
Report Date
August 21, 2014
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS BROKEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DOOR WAS BROKEN. THIS DOES NOT INDICATE A REPORTABLE MALFUNCTION. HOWEVER, DURING VERIFICATION TESTING AT THE SERVICE CENTER, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE DOOR ROLLER PIN WAS BROKEN.