RENEW ENDOCUT SCISSOR TIP, DISPOSABLE
Report
- Report Number
- 1223422-2014-00036
- Event Type
- Other
- Date Received
- October 17, 2014
- Report Date
- October 14, 2014
- Manufacturer
- MICROLINE SURGICAL INC
- Product Code
- GEI
- PMA / PMN Number
- K962119
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
INVESTIGATION FINDINGS: THE SCISSOR TIP WAS RETURNED AND INVESTIGATED. CUSTOMER'S STATED COMPLAINT THAT DURING A SURGICAL PT. THE SCISSOR TIP WAS VISUALLY INSPECTED AND MECHANICALLY TESTED AND THE SCISSOR TIP FUNCTIONED AS HANDPIECE THAT WAS RETURNED WITH THE SCISSOR TIP WAS INSPECTED. THE VISUAL AND MECHANICAL TESTING DID NOT REVEALED ANY NON-CONFORMING PERFORMANCE WITH THE DEVICE. THE SCISSOR TIP AND THE HANDPIECE WERE CONNECTED ACCORDING TO THE INSTRUCTIONS IN THE IFU. THE RESULT IS A TIGHT BOND BETWEEN THE TWO DEVICES. THE CONNECTION WAS TESTED WITH A PUSH AND PULL TEST. THESE TESTS WERE UNSUCCESSFUL FOR SEPARATING THE TWO DEVICES AND THE CUSTOMER'S STATED COMPLAINT COULD NOT BE DUPLICATED. HOWEVER, FAILURE TO FOLLOW THE IFU INSTRUCTIONS, MAY HAVE RESULTED IN AN INCORRECTLY INSTALLED SCISSOR TIP WHICH WAS NOT COMPLETELY TIGHTENED ONTO THE HANDPIECE AND MAY HAVE BECOME LOOSE.
DURING A SURGICAL PROCEDURE, THE SCISSOR TIP FELL OFF THE HANDLE AND INTO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 661429 | RENEW ENDOCUT SCISSOR TIP, DISPOSABLE | MANUAL DETACHABLE SURGICAL INSTRUMENTS | GEI | MICROLINE SURGICAL INC | 3142 | 00117864 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |