FDA Adverse Event Other Summary report: N

RENEW ENDOCUT SCISSOR TIP, DISPOSABLE

MDR report key: 4188452 · Received October 17, 2014

Report

Report Number
1223422-2014-00036
Event Type
Other
Date Received
October 17, 2014
Report Date
October 14, 2014
Manufacturer
MICROLINE SURGICAL INC
Product Code
GEI
PMA / PMN Number
K962119
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FINDINGS: THE SCISSOR TIP WAS RETURNED AND INVESTIGATED. CUSTOMER'S STATED COMPLAINT THAT DURING A SURGICAL PT. THE SCISSOR TIP WAS VISUALLY INSPECTED AND MECHANICALLY TESTED AND THE SCISSOR TIP FUNCTIONED AS HANDPIECE THAT WAS RETURNED WITH THE SCISSOR TIP WAS INSPECTED. THE VISUAL AND MECHANICAL TESTING DID NOT REVEALED ANY NON-CONFORMING PERFORMANCE WITH THE DEVICE. THE SCISSOR TIP AND THE HANDPIECE WERE CONNECTED ACCORDING TO THE INSTRUCTIONS IN THE IFU. THE RESULT IS A TIGHT BOND BETWEEN THE TWO DEVICES. THE CONNECTION WAS TESTED WITH A PUSH AND PULL TEST. THESE TESTS WERE UNSUCCESSFUL FOR SEPARATING THE TWO DEVICES AND THE CUSTOMER'S STATED COMPLAINT COULD NOT BE DUPLICATED. HOWEVER, FAILURE TO FOLLOW THE IFU INSTRUCTIONS, MAY HAVE RESULTED IN AN INCORRECTLY INSTALLED SCISSOR TIP WHICH WAS NOT COMPLETELY TIGHTENED ONTO THE HANDPIECE AND MAY HAVE BECOME LOOSE.

Description of Event or Problem · 1

DURING A SURGICAL PROCEDURE, THE SCISSOR TIP FELL OFF THE HANDLE AND INTO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661429 RENEW ENDOCUT SCISSOR TIP, DISPOSABLE MANUAL DETACHABLE SURGICAL INSTRUMENTS GEI MICROLINE SURGICAL INC 3142 00117864

Patients

Seq Age Sex Outcome Treatment
1 Other