FDA Adverse Event Injury Summary report: N

AIRFLOW RESUSCITATION BAG

MDR report key: 4188118 · Received September 19, 2014

Report

Report Number
2246980-2014-00017
Event Type
Injury
Date Received
September 19, 2014
Date of Event
September 5, 2013
Report Date
September 18, 2014
Manufacturer
VENTLAB
Product Code
BTM
PMA / PMN Number
K012842
Removal / Correction Number
RES68242
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS BELATED MDR IS BEING SUBMITTED AS PART OF OUR REMEDIATION REPORTS IN RESPONSE TO THE FDA FORM 483 NOTIFICATION ISSUED ON JUNE 4, 2014 BY (B)(4) TO VENTLAB, LLC. THIS REPORT CONSISTS OF 48 MDR'S. IT WAS CONFIRMED THAT THE BANG WOULD NOT DEFLATE BECAUSE THE PT VALVE HAD A STUCK DUCKBILL. THE DUCKBILL BECAME UNSTUCK DURING THE INVESTIGATION BUT IT TOOK 80CMH2O OF PRESSURE TO OPEN THE VALVE. ADDITIONALLY, VENTLAB PERFORMED QUALITY TRENDING, PRODUCT INVESTIGATION AND CORRECTIVE ACTIONS TO SOLVE THE ROOT CAUSES. THE SOLUTIONS WERE IMPLEMENTED LAST YEAR AND A RECALL WAS INITIATED IN MAY OF THIS YEAR.

Description of Event or Problem · 1

DURING PT USE THE BAG WOULD NOT DEFLATE. NO SERIOUS INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581084 AIRFLOW RESUSCITATION BAG MANUAL RESUSCITATOR BTM VENTLAB AF5140MB 200221

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention