AIRFLOW RESUSCITATION BAG
Report
- Report Number
- 2246980-2014-00008
- Event Type
- Injury
- Date Received
- September 19, 2014
- Date of Event
- April 29, 2013
- Report Date
- September 18, 2014
- Manufacturer
- VENTLAB
- Product Code
- BTM
- PMA / PMN Number
- K012842
- Removal / Correction Number
- RES68242
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS BELATED MDR IS BEING SUBMITTED AS PART OF OUR REMEDIATION REPORTS IN RESPONSE TO THE FDA FORM 483 NOTIFICATION ISSUED ON JUNE 4, 2014 BY (B)(4) TO VENTLAB, LLC. THIS REPORT CONSISTS OF 48 MDR'S. CUSTOMER RETURNED ALL PRODUCT FOR INVESTIGATION AND MALFUNCTION WAS CONFIRMED. BREAKDOWN: (B)(4) UNITS OPENED IMMEDIATELY, (B)(4) OPENED WITH LESS THAN 40CMH2O PRESSURE, AND (B)(4) OPENED WITH MORE THAN 40CMH2O OF PRESSURE. ADDITIONALLY, VENTLAB PERFORMED QUALITY TRENDING, PRODUCT INVESTIGATION AND CORRECTIVE ACTIONS TO SOLVE THE ROOT CAUSES. THE SOLUTIONS WERE IMPLEMENTED LAST YEAR AND A RECALL WAS INITIATED IN MAY OF THIS YEAR.
THE DUCKBILL VALVE INSIDE THE PT VALVE IS STUCK UPON FIRST INITIAL USE. VALVE CAN USUALLY BE FORCEFULLY BLOWN OPEN BY A HARD SQUEEZE OF BAG. ONCE OPEN IT APPEARS TO FUNCTION PROPERLY. SOME VALVES ARE HARDER TO OPEN THAN OTHERS. REPORTED NO INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 581087 | AIRFLOW RESUSCITATION BAG | MANUAL RESUSCITATOR | BTM | VENTLAB | AF1140MB-P | 105579, 106003, 106564 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |