FDA Adverse Event Injury Summary report: N

AIRFLOW RESUSCITATION BAG

MDR report key: 4188101 · Received September 19, 2014

Report

Report Number
2246980-2014-00008
Event Type
Injury
Date Received
September 19, 2014
Date of Event
April 29, 2013
Report Date
September 18, 2014
Manufacturer
VENTLAB
Product Code
BTM
PMA / PMN Number
K012842
Removal / Correction Number
RES68242
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS BELATED MDR IS BEING SUBMITTED AS PART OF OUR REMEDIATION REPORTS IN RESPONSE TO THE FDA FORM 483 NOTIFICATION ISSUED ON JUNE 4, 2014 BY (B)(4) TO VENTLAB, LLC. THIS REPORT CONSISTS OF 48 MDR'S. CUSTOMER RETURNED ALL PRODUCT FOR INVESTIGATION AND MALFUNCTION WAS CONFIRMED. BREAKDOWN: (B)(4) UNITS OPENED IMMEDIATELY, (B)(4) OPENED WITH LESS THAN 40CMH2O PRESSURE, AND (B)(4) OPENED WITH MORE THAN 40CMH2O OF PRESSURE. ADDITIONALLY, VENTLAB PERFORMED QUALITY TRENDING, PRODUCT INVESTIGATION AND CORRECTIVE ACTIONS TO SOLVE THE ROOT CAUSES. THE SOLUTIONS WERE IMPLEMENTED LAST YEAR AND A RECALL WAS INITIATED IN MAY OF THIS YEAR.

Description of Event or Problem · 1

THE DUCKBILL VALVE INSIDE THE PT VALVE IS STUCK UPON FIRST INITIAL USE. VALVE CAN USUALLY BE FORCEFULLY BLOWN OPEN BY A HARD SQUEEZE OF BAG. ONCE OPEN IT APPEARS TO FUNCTION PROPERLY. SOME VALVES ARE HARDER TO OPEN THAN OTHERS. REPORTED NO INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581087 AIRFLOW RESUSCITATION BAG MANUAL RESUSCITATOR BTM VENTLAB AF1140MB-P 105579, 106003, 106564

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention