FDA Adverse Event Injury Summary report: N

AVI

MDR report key: 41864 · Received September 13, 1996

Report

Report Number
41864
Event Type
Injury
Date Received
September 13, 1996
Date of Event
August 5, 1996
Report Date
August 13, 1996
Manufacturer
3M AVI, INC.
Product Code
FRN
Product Problem
Yes
Report Source
User Facility report
Reporter Location
OK, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PT DISCOVERED IN VENTRICULAR TACHYCARDIA. STAFF ON INVESTIGATION, FOUND PUMP INFUSING POTASSIUM PIGGYBACK AT 800+CC/HR. UNKNOWN HOW MUCH POTASSIUM HAD INFUSED AT THAT RATE. PT WAS HYPOTENSIVE BUT RESPONDED POSITIVELY TO CALCIUM CHLORIDE AND REVERTED TO SINUS RHYTHM. BLOOD PRESSURE NORMALIZED. THE PUMP WAS SET WITH AN NS FLUSH BAG ON THE MAIN LINE AND THE POTASSIUM PIGGYBACKED INTO THE NS. THE NURSE WHO HUNG THE PIGGYBACK OBTAINED A NEW PUMP. IT WAS PLUGGED IN. HE STRUNG THE NS AS THE PRIMARY IV BUT DID NOT SET A RATE. THE PIGGYBACK HAD ABOUT 80CC OF FLUID TO INFUSE AND WAS SET AT 60CC/HR. ORIGINAL PUMP BATTERY WAS FOUND TO BE DISCONNECTED INTERNALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AVI INFUSION PUMP FRN 3M AVI, INC. 480 *

Patients

Seq Age Sex Outcome Treatment
1 43 YR Required Intervention