FDA Adverse Event
Injury
Summary report: N
AVI
MDR report key: 41864
·
Received September 13, 1996
Report
- Report Number
- 41864
- Event Type
- Injury
- Date Received
- September 13, 1996
- Date of Event
- August 5, 1996
- Report Date
- August 13, 1996
- Manufacturer
- 3M AVI, INC.
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- OK, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
PT DISCOVERED IN VENTRICULAR TACHYCARDIA. STAFF ON INVESTIGATION, FOUND PUMP INFUSING POTASSIUM PIGGYBACK AT 800+CC/HR. UNKNOWN HOW MUCH POTASSIUM HAD INFUSED AT THAT RATE. PT WAS HYPOTENSIVE BUT RESPONDED POSITIVELY TO CALCIUM CHLORIDE AND REVERTED TO SINUS RHYTHM. BLOOD PRESSURE NORMALIZED. THE PUMP WAS SET WITH AN NS FLUSH BAG ON THE MAIN LINE AND THE POTASSIUM PIGGYBACKED INTO THE NS. THE NURSE WHO HUNG THE PIGGYBACK OBTAINED A NEW PUMP. IT WAS PLUGGED IN. HE STRUNG THE NS AS THE PRIMARY IV BUT DID NOT SET A RATE. THE PIGGYBACK HAD ABOUT 80CC OF FLUID TO INFUSE AND WAS SET AT 60CC/HR. ORIGINAL PUMP BATTERY WAS FOUND TO BE DISCONNECTED INTERNALLY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AVI | INFUSION PUMP | FRN | 3M AVI, INC. | 480 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Required Intervention |