FDA Adverse Event Malfunction Summary report: N

OT VERIO SYNC METER

MDR report key: 4183380 · Received October 18, 2014

Report

Report Number
2939301-2014-27875
Event Type
Malfunction
Date Received
October 18, 2014
Report Date
October 10, 2014
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K120708
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ¿ (02/09/2015). THE PATIENT¿S TEST STRIPS HAVE BEEN RETURNED ON 1/13/2015 AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 1/26/2015 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT FAILED TESTING. WHEN TEST STRIPS WERE TESTED WITH CONTROL SOLUTION, AN ERROR 4 WAS OBSERVED WITH NO ASSIGNABLE CAUSE FOUND. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 3 ¿ (03/18/2015) CORRECTION.SUPPLEMENTAL SENT TO CORRECT ALERT DATE. ALERT DATE SHOULD HAVE STATED 03/10/2015.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 - 3/18/2015 DEVICE EVALUATION.THE LAY USER/PATIENTS METER HAS BEEN RETURNED ON 2/27/2015 AND FURTHER EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON 3/10/2015 WITH THE FOLLOWING FINDINGS:ERROR MESSAGE 4 OBSERVED IN THE ERROR LOG, BUT THE ERRORS WERE NOT REPRODUCED DURING THE INVESTIGATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING ERROR 4. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
663896 OT VERIO SYNC METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3645791

Patients

Seq Age Sex Outcome Treatment
1 43 YR