ABBOTT M2000SP
Report
- Report Number
- 3005248192-2014-00065
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 24, 2014
- Report Date
- September 24, 2014
- Manufacturer
- ABBOTT MOLECULAR, INC.
- Product Code
- JJH
- PMA / PMN Number
- K092705
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
THE ELEVATED COMPLAINT INVESTIGATION (ECINV) FOR (B)(4) DETERMINED THIS COMPLAINT IS THE SAME ISSUE AS THAT OF (B)(4). SUMMARY OF ELEVATED COMPLAINT INVESTIGATION (B)(4): PROBLEM: EXCESSIVE NOISE FROM THE LIHA AND ROMA X-DRIVE MOTORS WAS REPORTED. NOISE AND STEP-LOSS FAILURES OCCUR WITHIN 2-3 MONTHS OF INSTRUMENT INSTALLATION OR REPLACEMENT SPARE PART INSTALLATION. CONCLUSIONS: RETAIN TESTING OF THE X-DRIVE MOTOR WAS NOT APPLICABLE AS NO FILE SAMPLES ARE RETAINED. SUPPLIER (B)(4) HAS CONFIRMED A PRODUCT DEFICIENCY WITH THE DRIVE BELT PRODUCTION PROCESS ON SOME BELTS. WITHIN 2-3 MONTHS OF OPERATION THE BELT TENSION CHANGES LEADING TO EXCESSIVE NOISE AND IN SOME CASES THE ASSOCIATED ARM CRASHES INTO A WORKTABLE OBJECT OR BECOMES MISALIGNED. X-DRIVE AXIS MOTOR REPLACEMENT SPARE PARTS AND ALL M2000SP AND M24SP INSTRUMENTS MANUFACTURED WITH THE ASSEMBLIES WITH DATE CODES BETWEEN 01 JANUARY 2013 THROUGH 30 SEPTEMBER 2013 ARE SUSPECT. THE M2000SP SERVICE MANUAL CONTAINS ADEQUATE INSTRUCTIONS FOR TROUBLESHOOTING X-DRIVE MOTOR ISSUES WITH REPLACEMENT INSTRUCTIONS AND PART NUMBER INFORMATION. VALID MATERIAL SPECIFICATIONS ARE RESIDENT IN PILGRIM SMARTSOLVE FOR THE MOTOR DC ARM X-AXIS 1000 INCREMENTS EVO PART NUMBER 10649045. A THOROUGH COMPLAINT SEARCH WAS PERFORMED TO IDENTIFY PAST COMPLAINT TICKETS FROM THE LAST TWO YEARS FOR ANY INSTANCES OF THE X-DRIVE MOTORS MAKING EXCESSIVE NOISE ON THE M2000SP INSTRUMENT. COMPLAINT HISTORY REVIEW SHOWED THAT, OVER THE LAST TWO YEARS, THIRTEEN COMPLAINTS INCLUDING THE SUSPECT COMPLAINTS ((B)(4)) WERE FOUND TO BE RELATED TO THE ISSUE OF THE X-DRIVE MOTOR FAILING ON THE M2000SP. THEREFORE, AS OF (B)(4) 2013 THERE WAS AN OCCURRENCE RATE OF THIRTEEN COMPLAINTS OVER THE LAST TWO YEARS. ELEVATED COMPLAINT INVESTIGATION (B)(4) CONCLUDED THERE WAS A CONFIRMED PRODUCT DEFICIENCY WITH THE X-DRIVE MOTOR PART NUMBER 10649045. ABBOTT MOLECULAR DETERMINED A FIELD ACTION (CORRECTION) SHOULD BE TAKEN TO INSPECT AND, IF NECESSARY, REPLACE THE AFFECTED BELTS. ABBOTT MOLECULAR INITIATED FIELD ACTION NUMBER (B)(4) ON NOVEMBER 23, 2013 AS A RECORDABLE FIELD ACTION. UPON RECEIPT OF FDA 483 OBSERVATION #1 ON JULY 11, 2014, ABBOTT MOLECULAR DETERMINED THIS FIELD ACTION SHOULD HAVE BEEN REPORTED TO FDA PER 21 CFR 806, AND ON (B)(4) 2014 REPORTED IT TO THE FDA (B)(4). TO DATE, AN FDA RECALL NUMBER HAS NOT BEEN RECEIVED. IN ADDITION TO MDR 3005248192-2014-00065, THE FOLLOWING MDRS WERE ALSO REPORTED: MDR 3005248192-2014-00007, MDR 3005248192-2014-00008, MDR 3005248192-2014-00009, MDR 3005248192-2014-00010, MDR 3005248192-2014-00011, MDR 3005248192-2014-00012, MDR 3005248192-2014-00013, MDR 3005248192-2014-00014, MDR 3005248192-2014-00015, MDR 3005248192-2014-00016, MDR 3005248192-2014-00017, MDR 3005248192-2014-00018, MDR 3005248192-2014-00019, MDR 3005248192-2014-00020, MDR 3005248192-2014-00021, MDR 3005248192-2014-00022, MDR 3005248192-2014-00023, MDR 3005248192-2014-00024, MDR 3005248192-2014-00025, MDR 3005248192-2014-00026, MDR 3005248192-2014-00027, MDR 3005248192-2014-00028, MDR 3005248192-2014-00029, MDR 3005248192-2014-00030, MDR 3005248192-2014-00031, MDR 3005248192-2014-00032, MDR 3005248192-2014-00033, MDR 3005248192-2014-00034, MDR 3005248192-2014-00035 MDR 3005248192-2014-00036, MDR 3005248192-2014-00037, MDR 3005248192-2014-00038, MDR 3005248192-2014-00039, MDR 3005248192-2014-00040, MDR 3005248192-2014-00041, MDR 3005248192-2014-00042, MDR 3005248192-2014-00043, MDR 3005248192-2014-00044, MDR 3005248192-2014-00045, MDR 3005248192-2014-00046, MDR 3005248192-2014-00047, MDR 3005248192-2014-00048, MDR 3005248192-2014-00049, MDR 3005248192-2014-00050, MDR 3005248192-2014-00051, MDR 3005248192-2014-00052, MDR 3005248192-2014-00053, MDR 3005248192-2014-00054, MDR 3005248192-2014-00055, MDR 3005248192-2014-00058, MDR 3005248192-2014-00059, MDR 3005248192-2014-00060, MDR 3005248192-2014-00063, MDR 3005248192-2014-00064.
THE ABBOTT M2000 SYSTEM IS INTENDED FOR USE IN PERFORMING NUCLEIC ACID AMPLIFICATION TESTING BY POLYMERASE CHAIN REACTION IN CLINICAL LABORATORIES. THE SYSTEM IS COMPOSED OF THE M2000RT AND THE M2000SP INSTRUMENTS. THE M2000SP IS INTENDED FOR AUTOMATED SAMPLE PREPARATION AND NUCLEIC ACID EXTRACTION PRIOR TO TESTING. THE INSTRUMENT INVOLVED IN THIS COMPLAINT IS AN ABBOTT M24SP INSTRUMENT (LIST 03N06-01), SERIAL NUMBER (B)(4).THE ABBOTT M24SP IS A CE-MARKED EXTRACTION INSTRUMENT NOT SOLD IN THE UNITED STATES; THIS INCIDENT IS REPORTED ON THE BASIS THAT THE M24SP INSTRUMENT IS SIMILAR TO THE ABBOTT M2000SP INSTRUMENT (LIST 09K14-02) SOLD IN THE UNITED STATES. THE LIQUID HANDLING ARM (LIHA) IS A MECHANICAL COMPONENT OF THE M24SP INSTRUMENT THAT PIPETTES AND DISPENSES SAMPLES AND REAGENTS. A CUSTOMER REPORTED THAT THE LIHA COMPONENT CRASHED AND THAT THE M24SP COULD NOT BE INITIALIZED. THE FIELD SERVICE ENGINEER (FSE) FOUND THAT THE X BELT WAS BROKEN AND DISCOLORED. AS THE SYMPTOMS WERE POTENTIALLY RELEVANT TO A KNOWN X-AXIS BELT FAILURE IDENTIFIED IN (B)(4) "M24SP X-DRIVE ASSEMBLY BELT INSPECTION AND REPLACEMENT PROCEDURE", THE FSE EXECUTED THE TSB AND REPLACED THE BELT. THE INSTRUMENT RAN SMOOTHLY WHITHOUT ANY ERROR. THERE WAS NOT ANY REPORT OF DEATH OR INJURY. ELEVATED COMPLAINT INVESTIGATION (B)(4) CONFIRMED THAT THIS COMPLAINT IS LINKED TO A KNOWN PRODUCT MALFUNCTION OF THE X-DRIVE MOTOR. SUMMARY OF ORIGINAL ELEVATED COMPLAINT INVESTIGATION (B)(4). MEDICAL EVALUATION CONCLUDED THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO DEATH OR SERIOUS INJURY. THIS COMPLAINT IS REPORTED IN ACCORDANCE WITH 21 CFR 803 ON THE BASIS IT IS LINKED TO REPORTABLE (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 660915 | ABBOTT M2000SP | CLINICAL SAMPLE CONCENTRATOR | JJH | ABBOTT MOLECULAR, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |