FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 4183290 · Received October 17, 2014

Report

Report Number
2531779-2014-29674
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
October 7, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 12/09/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 11/20/2014 WITH THE FOLLOWING FINDINGS: DUE TO MOISTURE CONTAMINATION AND BATTERY COMPARTMENT CRACKS, THE PUMP INTERMITTENTLY POWERED ON WITH THE RETURNED BATTERY CAP. THE BATTERY CAP WAS UNABLE TO FULLY TIGHTEN. BATTERY COMPARTMENT CRACKS WERE OBSERVED IN THE THREAD AREA AND BELOW THE GRIP PAD. THERE WAS EVIDENCE OF MOISTURE CONTAMINATION INSIDE THE BATTERY COMPARTMENT. PUMP REBOOT EVENTS WERE OBSERVED IN THE BLACK BOX ON 10/7/2014. A LEAK TEST SHOWED THAT THERE WAS A LEAK AT A BATTERY COMPARTMENT CRACK. THERE WAS EVIDENCE OF ADDITIONAL MOISTURE INSIDE THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE PUMP WOULD NOT POWER ON. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660143 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR