PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4183224 · Received October 17, 2014

Report

Report Number
2032227-2014-39779
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 19, 2014
Report Date
September 19, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH ALL BUTTONS FUNCTIONING PROPERLY. HOWEVER CORRODED KEYPAD TRACES NOTED. NO BUTTON ERROR ALARMS NOTED. THE INSULIN PUMP HAS CRACKED RESERVOIR TUBE LIP, CRACKED BELT CLIP SLOT, MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. IT WAS STATED THAT THE ALARM IS RECURRING. IT WAS STATED THAT THE LOGO IS DISCOLORED. AN OPERATION TEST AND A SELF TEST WERE COMPLETED NORMALLY. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660576 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LWWL

Patients

Seq Age Sex Outcome Treatment
1