FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4183195 · Received October 17, 2014

Report

Report Number
2032227-2014-40075
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 19, 2014
Report Date
September 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Report Source
Manufacturer report
Reporter Location
GA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.(B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND DISPLACEMENT TESTS. NO PRIME FILL ANOMALY NOTED. THE INSULIN PUMP WAS RECEIVED WITH BROKEN BELT CLIP SLOT AND MINOR SCRATCHES ON LCD WINDOW.

Description of Event or Problem · 1

NO BLOOD GLUCOSE LEVELS REPORTED. THE CUSTOMER REPORTED BEING UNABLE TO GET OUT OF THE FILLING SCREEN ON THE INSULIN PUMP. THE CUSTOMER EXPLAINED THAT THEY WERE UNABLE TO EXIT THE PRIME LOOP FOR THE INSULIN PUMP. TROUBLESHOOTING WAS DONE. THE CUSTOMER WAS INSTRUCTED ON HOW TO EXIT THE PRIME LOOP. THE INSULIN PUMP DID NOT EXIT THE PRIME REWIND LOOP OF THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660339 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 57 YR