8 CM ANGLE ATTACHMENT
Report
- Report Number
- 1045834-2014-14351
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- August 14, 2012
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- PK011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- OTHER
Narratives
ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY TECHNICIAN, THE FUNCTIONAL TESTING REVEALED THAT DEVICE GETS HOT, ABOVE TEMPERATURE SPECIFICATIONS. IT WAS ALSO NOTED THAT THERE WAS DETERGENT RESIDUE AND RUST PRESENT. IMPROPER CLEANING AND MAINTENANCE WERE ATTRIBUTED TO THE REPORTED CONDITION. THE DEVICE WAS REPAIRED BY REPLACING SEVERAL COMPONENTS.
CUSTOMER REPORTED THE DEVICE DID NOT FUNCTION PROPERLY. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).