FDA Adverse Event Malfunction Summary report: N

10.5CM ANGLE ATTACHMENT, BLACK MAX

MDR report key: 4183160 · Received October 17, 2014

Report

Report Number
1045834-2014-14297
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
September 26, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK965080
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. THE DEVICE WAS RECEIVED FOR INSPECTION. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED BALL BEARINGS IN EXTENSION SLEEVE WERE RUSTED AND DAMAGED. THE DEVICE OVERHEATED AND FAILED VIBRATION TEST. THIS ATTRIBUTED TO IMPROPER MAINTENANCE. THE DEVICE WAS REPAIRED AND RETURNED TO CUSTOMER.

Description of Event or Problem · 1

THE CUSTOMER RETURNED THE DEVICE FOR SERVICE. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660652 10.5CM ANGLE ATTACHMENT, BLACK MAX DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1