FDA Adverse Event Malfunction Summary report: N

11CM ANGLE ATTACHMENT, BLACK MAX

MDR report key: 4183158 · Received October 17, 2014

Report

Report Number
1045834-2014-14155
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
July 22, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DEVICE WAS DAMAGED/OVERHEATED AND NOT FUNCTIONING PROPERLY. THIS WAS ATTRIBUTED TO NORMAL WEAR. DEVICE WAS REPLACED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR AN UNKNOWN REASON. DURING SERVICE, TECHNICIAN FOUND AXLE BEARING WORN FROM USE AND OVERHEATED. ALSO, RING COLOR AT THIMBLE HAS BEEN MISSING. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660320 11CM ANGLE ATTACHMENT, BLACK MAX DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1