FDA Adverse Event Malfunction Summary report: N

AO/ASIF-QC F/COMPACT AIR DRIVE+POWER DRI

MDR report key: 4183148 · Received October 17, 2014

Report

Report Number
8030965-2014-01562
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
January 29, 2013
Manufacturer
SYNTHES GMBH
Product Code
HWE
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. MANUFACTURING DATE FOR THIS DEVICE IS UNAVAILABLE DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE COUPLING SHAFT WAS BENT. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER ON (B)(4) 2103. PLACEHOLDER.

Description of Event or Problem · 1

CUSTOMER REPORTED THE A/O QUICK COUPLING SEEMS TO BE STUCK AND BENT. WHEN THE SURGEON USED IT THE DRILL HAD A WOBBLE IN IT. THIS IS REPORT 1 OF 1 FOR COMPLAINT#(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660581 AO/ASIF-QC F/COMPACT AIR DRIVE+POWER DRI HWE SYNTHES GMBH

Patients

Seq Age Sex Outcome Treatment
1