AO/ASIF-QC F/COMPACT AIR DRIVE+POWER DRI
Report
- Report Number
- 8030965-2014-01562
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- January 29, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. MANUFACTURING DATE FOR THIS DEVICE IS UNAVAILABLE DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE COUPLING SHAFT WAS BENT. THIS WAS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER ON (B)(4) 2103. PLACEHOLDER.
CUSTOMER REPORTED THE A/O QUICK COUPLING SEEMS TO BE STUCK AND BENT. WHEN THE SURGEON USED IT THE DRILL HAD A WOBBLE IN IT. THIS IS REPORT 1 OF 1 FOR COMPLAINT#(B)(4).