FDA Adverse Event Malfunction Summary report: N

LOW PRESSURE BLACK MAX HANDPIECE

MDR report key: 4183145 · Received October 17, 2014

Report

Report Number
1045834-2014-13944
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
January 28, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBB
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. MANUFACTURING DATE FOR THIS DEVICE IS UNAVAILABLE. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE RPM WERE BELOW SPECIFICATION. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE DEVICE DID NOT HAVE ENOUGH POWER. DURING SERVICE, TECHNICIAN INDICATED THE DEVICE WAS WORKING BUT IT HAD A LOW REVOLUTIONS PER MINUTE AND AN OLD HOSE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660574 LOW PRESSURE BLACK MAX HANDPIECE MOTOR, DRILL, PNEUMATIC HBB DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1