LOW PRESSURE BLACK MAX HANDPIECE
Report
- Report Number
- 1045834-2014-13944
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- January 28, 2013
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBB
- PMA / PMN Number
- PK831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. MANUFACTURING DATE FOR THIS DEVICE IS UNAVAILABLE. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE RPM WERE BELOW SPECIFICATION. THIS WAS ATTRIBUTED TO NORMAL WEAR. VARIOUS PARTS WERE REPLACED. DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.
CUSTOMER REPORTED THAT THE DEVICE DID NOT HAVE ENOUGH POWER. DURING SERVICE, TECHNICIAN INDICATED THE DEVICE WAS WORKING BUT IT HAD A LOW REVOLUTIONS PER MINUTE AND AN OLD HOSE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).