FDA Adverse Event Malfunction Summary report: N

BLACKMAX-NEURO LOW PRESSURE

MDR report key: 4183119 · Received October 17, 2014

Report

Report Number
1045834-2014-14080
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
August 19, 2013
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBB
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. THE DEVICE RETURNED FOR SERVICE DURING THE EVALUATION THE TECHNICIAN FOUND DAMAGED LOCKING COMPONENT. FAILED TEMPERATURE AND SPEED TESTS. THIS ATTRIBUTED TO NORMAL WEAR. THE DEVICE WAS SUBSEQUENTLY REPAIRED.

Description of Event or Problem · 1

DEVICE WAS RETURNED FOR SERVICE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661164 BLACKMAX-NEURO LOW PRESSURE MOTOR, DRILL, PNEUMATIC - HANDPIECE HBB DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1