FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4183114 · Received October 17, 2014

Report

Report Number
2032227-2014-40144
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 15, 2014
Report Date
September 19, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING BASIC OCCLUSION TEST DUE TO FAULTY FORCE SENSOR RELAY. THE MOTOR PASSED MOTOR TEST. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP AND CRACKED CASE AT RESERVOIR TUBE WINDOW CORNER.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING PRIME. DEVICE WAS NOT EXPOSED TO A MAGNETIC FIELD. CUSTOMER DOES NOT USE THE SENSOR FEATURE AND IS UNABLE TO COMPLETE THE REWIND SEQUENCE. BLOOD GLUCOSE VALUE IS 189 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660090 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522RNAS A000142707

Patients

Seq Age Sex Outcome Treatment
1 74 YR