FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 4183083 · Received October 17, 2014

Report

Report Number
2032227-2014-40065
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
September 19, 2014
Report Date
September 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED SELF TEST AND DISPLACEMENT TEST. NO UNEXPECTED VIBRATION ANOMALY NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON LCD WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS AND BATTERY TUBE THREADS.

Additional Manufacturer Narrative · 1

(B)(4). THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP WAS NOT VIBRATING. CUSTOMER STATED THAT THE ANOMALY WAS PERSISTING. SELF TEST WAS PERFORMED AND PASSED. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660479 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-754WWB

Patients

Seq Age Sex Outcome Treatment
1