FDA Adverse Event Malfunction Summary report: N

V60 VENTILATOR

MDR report key: 4182845 · Received October 17, 2014

Report

Report Number
2031642-2014-01252
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
September 26, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
MNT
PMA / PMN Number
K082660
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Description of Event or Problem · 1

THE INTERNATIONAL SERVICE TECH REPORTED THAT DURING SERVICING OF THE UNIT THE BATTERY NOT CHARGING. POWER FAILURE DUE TO LOSS OF BATTERY POWER MAY RESULT IN SHUTDOWN OF DEVICE DURING NORMAL VENTILATION OPERATION. THE INTERNATIONAL SERVICE TECH REPLACED THE BATTERY AND THE REPORTED PROBLEM RESOLVED. THE UNIT PASSED ALL APPLICABLE TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661370 V60 VENTILATOR VENTILATOR, CONTINUOUS MNT RESPIRONICS CALIFORNIA, INC. V60 NA

Patients

Seq Age Sex Outcome Treatment
1