530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-39958
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 18, 2014
- Report Date
- September 18, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. (B)(4)
CUSTOMER REPORTS TO HAVE RECEIVED A MOTOR ERROR ALARM. CUSTOMER'S BLOOD GLUCOSE WAS BETWEEN 140 MG/DL AND 150 MG/DL. DURING TROUBLESHOOTING, CUSTOMER WAS ABLE TO COMPLETE THE REWIND SEQUENCE. AFTER TROUBLESHOOTING, CUSTOMER WAS ADVISED THAT INSULIN PUMP WOULD NEED TO BE REPLACED. NO FURTHER INFORMATION PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 661921 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-551NAL |