Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND AN EVALUATION WAS COMPLETED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT. THE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED IN THE EVENT HISTORY LOG REVIEW. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. UPON CONCLUSION OF THE INVESTIGATION, IT WAS DETERMINED THAT THE CAUSE OF THE IIPV EVENT WAS DUE TO A FALSE EMPTY DETECT AND USE ERROR, FURTHER SPECIFIED AS THE INITIAL DRAIN ALARM SETTING WAS INAPPROPRIATELY PROGRAMMED. THE HOMECHOICE / HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE WARNS THAT ¿SETTING THE I-DRAIN ALARM VOLUME TOO LOW OR OFF CAN RESULT IN AN INCOMPLETE INITIAL DRAIN FOLLOWED BY A FULL FILL. THIS CAN RESULT IN AN INCREASED INTRAPERITONEAL VOLUME (IIPV) SITUATION.¿ A REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD RELEVANT ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED. THIS EVENT OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 AT 22:05:53. DURING NIGHT DRAIN CYCLE 1, THE PATIENT'S ULTRAFILTRATION READING WAS 927ML, INDICATING THE HOME PATIENT DRAINED 927ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 661919 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |