PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-39966
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 19, 2014
- Report Date
- September 19, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP ALARMED A33 ERROR DURING BASIC OCCLUSION TEST AND UNABLE TO PRIME AT 4.0 LBS DURING PRIME/A33 TEST DUE TO PROTRUDED LOOSE DRIVE SUPPORT DISK. THE INSULIN PUMP HAS CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.
NO BLOOD GLUCOSE LEVELS REPORTED. THE CUSTOMER REPORTED BEING UNABLE TO GET OUT OF THE FILLING SCREEN ON THE INSULIN PUMP. THE CUSTOMER EXPLAINED THAT THEY WERE UNABLE TO EXIT THE PRIME LOOP FOR THE INSULIN PUMP. TROUBLESHOOTING WAS DONE. THE CUSTOMER WAS INSTRUCTED ON HOW TO EXIT THE PRIME LOOP. THE INSULIN PUMP DID NOT EXIT THE PRIME REWIND LOOP OF THE INSULIN PUMP. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WOULD NEED TO BE REPLACED. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO A BACK-UP PLAN PER THE HEALTH CARE PROVIDER'S INSTRUCTION. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 661916 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 44 YR |