FDA Adverse Event Malfunction Summary report: N

THE EMAX 2 PLUS SYSTEM

MDR report key: 4182720 · Received October 17, 2014

Report

Report Number
1045834-2014-14651
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
December 12, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES OUS SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DURING PRE-REPAIR ASSESSMENT PERFORMED BY A TECHNICIAN, FUNCTIONAL TESTING WAS PERFORMED AND REVEALED THE DEVICE WAS FILLED WITH OIL AND HAD LUBRICATION DAMAGE, THE BEARING WAS FAULTY, THE MOTOR AND MOTOR FLEX WERE FAULTY, AND THE MOTOR OVERHEATED. THIS WAS ATTRIBUTED TO DEFECTIVE CLEANING AND MAINTENANCE. THE DEVICE WAS REPAIRED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT DRILL HEATED UP DURING USE. THIS IS REPORT 1 OF 1 FOR (B)(4).