530G INSULIN PUMP
Report
- Report Number
- 2032227-2014-39876
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Date of Event
- September 19, 2014
- Report Date
- September 19, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE. (B)(4).
IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP MAY NOT BE FUNCTIONING PROPERLY DUE TO HIGH BLOOD GLUCOSE LEVELS. CUSTOMER STATED THEY HAD HIGH BLOOD GLUCOSE LEVELS OF 440 MG/DL AND ARE TREATING WITH MANUAL INSULIN INJECTIONS. CUSTOMER ADVISED TO DISCONNECT FROM PUMP. DURING TROUBLESHOOTING, CUSTOMER WAS ASKED TO CHECK THE CONDITION OF THE DRIVE SUPPORT CAP ON THE DEVICE AND FOUND IT TO BE NORMAL. NEXT, CUSTOMER WAS ASKED TO CHECK FOR AIR BUBBLES IN INFUSION SET TUBING. CUSTOMER VERIFIED THERE WERE NO AIR BUBBLES. AFTERWARDS, THE RESERVOIR WAS REINSERTED INTO INSULIN PUMP AND A MANUAL PRIME WAS PERFORMED. DEVICE PASSED PRIMING BUT DEVICE ALARMED MOTOR ERROR DURING HIGH PRESSURE TESTING. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 661021 | 530G INSULIN PUMP | OZO | OZO | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-551NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |