FDA Adverse Event Injury Summary report: N

530G INSULIN PUMP

MDR report key: 4182636 · Received October 17, 2014

Report

Report Number
2032227-2014-39713
Event Type
Injury
Date Received
October 17, 2014
Date of Event
September 19, 2014
Report Date
September 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZO
PMA / PMN Number
120010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED (B)(4).

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACES. NO MOISTURE DAMAGE WAS NOTED ON THE ELECTRONICS. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT THE DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE INSULIN PUMP BUTTONS WERE UNRESPONSIVE. TROUBLESHOOTING WAS DONE. CUSTOMER'S BLOOD GLUCOSE WAS 427 MG/DL. CUSTOMER TREATED WITH MANUAL SYRINGE. CUSTOMER STATED THAT THE NUMBERS KEPT SCROLLING WHILE TRYING TO PROGRAM A BOLUS. ADVISED CUSTOMER THE INSULIN PUMP WOULD BE REPLACED. ADVISED TO DISCONTINUE USING THE PUMP AND REVERT TO A BACK-UP PLAN PER HEALTH CARE PROVIDER. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660958 530G INSULIN PUMP OZO OZO MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-751NAP

Patients

Seq Age Sex Outcome Treatment
1 34 YR