FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4182611 · Received October 17, 2014

Report

Report Number
2032227-2014-39867
Event Type
Malfunction
Date Received
October 17, 2014
Date of Event
August 30, 2014
Report Date
September 19, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Additional Manufacturer Narrative · 1

(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A MOTOR ERROR ALARM ON THE INSULIN PUMP AND STATED THAT THEY HAD HIGH BLOOD GLUCOSE READINGS DUE TO THE INSULIN PUMP NOT DELIVERING INSULIN. CUSTOMER STATED THAT THE INSULIN PUMP HAS BEEN THROUGH A BODY SCANNER AT THE AIRPORT. CUSTOMER DOES USE THE SENSOR FEATURE AND THEY WERE ABLE TO COMPLETE THE REWIND SEQUENCE. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE CALL WAS 250 MG/DL. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660436 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAB

Patients

Seq Age Sex Outcome Treatment
1 25 YR