FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 4182594 · Received October 17, 2014

Report

Report Number
2032227-2014-39696
Event Type
Injury
Date Received
October 17, 2014
Date of Event
September 9, 2014
Report Date
September 30, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO FLATTENED ALL THE BUTTONS DOME SWITCH. NO STUCK BUTTONS NOTED. UNIT HAD NORMAL OPERATING CURRENTS AND NO UNEXPECTED OFF NO POWER OR LOW BATTERY ALARM NOTED. UNIT PASSED FUNCTIONAL TESTINGS INCLUDING DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND NO DELIVERY TESTS. UNIT HAD MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP AND STAINED END CAP STICKER. NO BURNT MARKS NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

THE CUSTOMER REPORTED HOSPITALIZATION AND THAT THE INSULIN BUTTONS WERE UNRESPONSIVE. CUSTOMER STATED THAT THE BUTTONS WERE STICKING. TROUBLESHOOTING WAS DONE. CUSTOMER'S BLOOD GLUCOSE WAS 123 MG/DL. CUSTOMER MENTIONED THAT THE INSULIN PUMP LOOKS BURNT. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. CUSTOMER STATED ALSO THAT THERE WAS EMERGENCY ROOM VISIT DUE TO HIGH BLOOD GLUCOSE. CUSTOMER'S BLOOD GLUCOSE WAS 400 MG/DL. THE CAUSE OF THE EMERGENCY VISIT WAS UNCONTROLLED DIABETES. CUSTOMER STAYED FOR A FEW DAYS AND WAS RELEASED. CUSTOMER WAS WEARING THE INSULIN PUMP AT THE TIME OF EMERGENCY ROOM VISIT. NO FURTHER INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660256 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAH

Patients

Seq Age Sex Outcome Treatment
1 39 YR Hospitalization