FDA Adverse Event Injury Summary report: N

SHAILEY TRACH 8DCT

MDR report key: 418255 · Received September 19, 2002

Report

Report Number
418255
Event Type
Injury
Date Received
September 19, 2002
Date of Event
September 17, 2002
Report Date
September 18, 2002
Manufacturer
MALLINCKRODT
Product Code
JOH
Report Source
User Facility report
Reporter Location
MA, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

VENT ALARM FOR LOW EXHALED MINUTE VOLUME AND LOW EXHALED TIDAL VOLUME. AUDIBLE CUFF LEAK, INABILITY BTO MAINTAIN PATENCY OF THE CUFF. THE PILOT BALLOON WAS PLACED IN WATER AND NOTED AIR BUBBLES ALONG THE HEAT SEAL EDGE OF THE BALLOON. PILOT BALLOON CRIMPED OFF AND NEW SHAILEY TRACH PLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SHAILEY TRACH 8DCT TRACH TUBE 8DCT JOH MALLINCKRODT * *

Patients

Seq Age Sex Outcome Treatment
1 64 YR Hospitalization| L 603-881-5296