FDA Adverse Event Malfunction Summary report: N

8CM ANGLE ATTACHMENT, BLACK MAX

MDR report key: 4182533 · Received October 17, 2014

Report

Report Number
1045834-2014-13946
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
August 17, 2012
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBE
PMA / PMN Number
PK831756
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND DAMAGED BEARINGS WAS FOUND NORMAL WEAR. BEARING IS DEFAULT, IT WAS HOT. BEARINGS BALL WAS RUSTY. COUPLING AND BEARING WERE REPLACED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE IS HEATING. REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
660501 8CM ANGLE ATTACHMENT, BLACK MAX DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1