FDA Adverse Event
Malfunction
Summary report: N
8CM ANGLE ATTACHMENT, BLACK MAX
MDR report key: 4182533
·
Received October 17, 2014
Report
- Report Number
- 1045834-2014-13946
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- August 17, 2012
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBE
- PMA / PMN Number
- PK831756
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BE
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ADDITIONAL NARRATIVE: SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO SYNTHES SERVICE AND REPAIR FILES LEGACY REVIEW/REMEDIATION PROTOCOL-COMPLAINT HANDLING AND MDR REPORTING. DEVICE LISTED IN THIS REPORT IS USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. DEVICE WAS RETURNED FOR SERVICE. DURING SERVICE, TECHNICIAN FOUND DAMAGED BEARINGS WAS FOUND NORMAL WEAR. BEARING IS DEFAULT, IT WAS HOT. BEARINGS BALL WAS RUSTY. COUPLING AND BEARING WERE REPLACED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DEVICE IS HEATING. REPORT 1 OF 1 FOR (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 660501 | 8CM ANGLE ATTACHMENT, BLACK MAX | DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) | HBE | DEPUY SYNTHES POWER TOOLS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |