FDA Adverse Event Malfunction Summary report: N

ANIMAS VIBE

MDR report key: 4182458 · Received October 17, 2014

Report

Report Number
2531779-2014-29618
Event Type
Malfunction
Date Received
October 17, 2014
Report Date
October 3, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 12/22/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. A REVIEW OF THE PUMP BLACK BOX DATA REVEALED CS 052 AND CS 054 CALL SERVICE ALARMS. DURING TESTING, THE PUMP WAS EXERCISED FOR 24 HOURS AND THE REWIND, LOAD, AND PRIME STEPS WERE COMPLETED SUCCESSFULLY WITH NO DUPLICATED ALARMS. THE PUMP CASE WAS REMOVED AND NO INTERNAL DEFECTS WERE FOUND.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CS 052/053/054/055 SLEEP) ISSUE. IT WAS REPORTED THAT THE PUMP EMITTED A CS 052 CALL SERVICE ALARM THREE TIMES IN THIRTY DAYS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
661252 ANIMAS VIBE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1