ANIMAS VIBE
Report
- Report Number
- 2531779-2014-29615
- Event Type
- Malfunction
- Date Received
- October 17, 2014
- Report Date
- October 5, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1: DATE OF SUBMISSION (B)(4) 2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE AUDIO BOLUS BUTTON WAS DIFFICULT TO PUSH AND WAS FOUND TO BE PUNCTURED/TORN. UNRELATED TO THE INITIAL COMPLAINT, A LINE WAS OBSERVED THROUGH THE DISPLAY. THE PUMP WAS OPENED AND CONTAMINATION ON THE FLEX RIBBON WAS OBSERVED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (ABB TACTILE CNG PRIOR TO DMG) ISSUE. THE REPORTER STATED THAT THE AUDIO BOLUS BUTTON WAS PEELING AND UNDER RESPONSIVE. NO ADDITIONAL INFORMATION WAS AVAILABLE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.