FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 4182418 · Received October 17, 2014

Report

Report Number
2024168-2014-06788
Event Type
Injury
Date Received
October 17, 2014
Date of Event
September 22, 2014
Report Date
September 22, 2014
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE IS EXPECTED TO BE RETURNED FOR EVALUATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). CORRECTION - IT WAS INITIALLY REPORTED THAT THE DEVICE WOULD BE RETURNED FOR ANALYSIS; HOWEVER, SUBSEQUENT INFORMATION REVEALED THAT THE DEVICE WAS DISCARDED AND IS NOT AVAILABLE FOR EVALUATION. IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE WITH A 6F SHEATH AFTER AN ENDOVASCULAR INTERVENTIONAL PROCEDURE. REPORTEDLY, A CUFF MISS OCCURRED. HEMOSTASIS WAS ACHIEVED USING A SECOND PROGLIDE DEVICE. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE WAS REPORTED. THE PHYSICIAN IS REPORTED TO BE TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
662146 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 40723K1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 6F